中国科技核心期刊
美国化学文摘社(CAS)数据库
美国EBSCO学术数据库
日本科学技术振兴机构数据库(JST)
The inductively coupled plasma-optical emission spectrometry was employed to measure the Pt in the platinum matrix solution during the testing of gold alloy jewelry.The interference in the platinum matrix solution of the Pt spectrum lines recommended in national standard GB/T 38145—2019 High Content Precious Metals Jewelry Alloys-Determination of Gold,Platinum and Palladium-Difference Method Using Inductively Coupled Plasma-Optical Emission Spectroscopy,Thermo IRIS Intrepid II inductively coupled plasma-optical emission spectrometer and PE Optima 8300 inductively coupled plasma-optical emission spectrometer is analyzed.On the basis of spectral line intensity,sensitivity,standard curve correlation coefficient,light intensity value and interference spectrum line position,the best spectrum line for practical testing the Pt element has been confirmed.