Abstract:
The samples were digested by aqua regia,with 3.0 mL iron trichloride added to the solution.After the solution being adsorbed to polyurethane foam plastics and then desorbed by thiourea,inductively coupled plasma mass spectrometry was proposed as the analytical method to measure the trace gold in geological samples.The results suggest that the measurement range of this method is 0.10 to 500.00 ng/g, the RSD is between 8.33 % and 9.01 %,and the recovery ranges from 94.0 % to 108.0 %.This method takes short time,is easy to operate and has low detection limit,high precision and accuracy.