Abstract:
The samples were digested with HCl-HNO3-HClO4 mixed acid.Then Cu,Pb,Zn and Cd in gold concentrates are determined in HCl media by inductively coupled plasma-atomic emission spectrometry.The detection limit is Cu 0.002 2 μg/mL,Pb 0.014 0 μg/mL,Zn 0.004 2 μg/mL,Cd 0.003 5 μg/mL respectively.The recovery rate of reference materials is Cu 97.67 %-102.25 %,Pb 96.97 %-105.07 %,Zn 97.00 %-105.50 %,Cd 95.00 %-105.00 %, respectively.The RSD is Cu 3.03 %-3.79 %,Pb 1.24 %-4.52 %,Zn 1.62 %-3.91 %,Cd 1.94 %-4.16 % respectively.The method is easy to operate,accurate and reliable,can meet the lab test requirement and is suitable for large batch test of samples.